Open original FDA record ↗Recall record
- FDA identifier
- Z-1569-2020
- Company in FDA record
- Skytron, Div. The KMW Group, Inc
- FDA record creation date
- Mar 25, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury.
Source facts
- Action
- Skytron customer notification letters mailed on 3/5/2020. Distributors received notification via email on 3/5/2020. Letter states reason for recall, health risk and action to take: 1.Identify and inspect your 3FCM1 monitor bracket(s). 1508T719-00-0-0120 2.Remove from service any bracket with noticeable physical distortion or cracking around the welded area. 3.Skytron authorized representatives have been directed to contact and schedule a service visit within a 30-day window of time. The contact information for your authorized representative is: Johnsons Medical, Inc. Donna Johnson, (801) 768-4411 Please contact Erin Woolf, at 616-656-2998 or ewoolf@skytron.us with any questions. For after-hours support, please call 1-800-SKYTRON.
- Initiated
- 2020-03-05
- Root cause
- Device Design
- Distribution
- AZ, CO, GA, IL, MI, MN, NJ, PA, VA, VT
- Recall number
- Z-1569-2020
- Recall status
- Open, Classified
- CDRH record ID
- 180367
- FDA record created
- 2020-03-25
- Recall event number
- 85252