Skip to content

Public FDA record

Z-1571-2022: Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vant

Recall record Z-1571-2022 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1571-2022
Company in FDA record
Medtronic Neuromodulation
FDA record creation date
Aug 24, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 to display an Unexpected Device Error Code 1502 Message, and the user will be unable to perform programming of the Vanta Implantable Neurostimulator (INS) Model 977006.

Source facts

Action
On about 07/26/2022, Medtronic notified consignees via mail courier service with a letter titled "URGENT: MEDICAL DEVICE CORRECTION." The letter instructed consignees to perform an initial interrogation of the Vanta INS and tap start usage, tap implant device workflow, click start, and on the next screen exit the session, and to initiate a second interrogation and if there is no "Unexpected Device Error Code 1502" code the INS maybe used. If the INS cannot be interrogated, contact Medtronic Technical Services at 1-800-707-0933. Additionally, consignees were instructed to complete and return the customer confirmation form provided.
Initiated
2022-07-26
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. The countries of Austria, Belgium, Canada, Canary Islands, Cayman Islands, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Kingdom.
Recall number
Z-1571-2022
Recall status
Open, Classified
CDRH record ID
195423
FDA record created
2022-08-24
Recall event number
90686

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗