Z-1572-2026: MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Recall record Z-1572-2026 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1572-2026
Company in FDA record
Howmedica Osteonics Corp.
FDA classification date
Mar 18, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Source facts

Action
On February 2, 2026 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and return via email. 4. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 5. Please return the devices to Stryker using the address provided. ***Update 04/22/2026*** On April 2 ,2026 MEDICAL DEVICE RECALL UPDATE letters were sent to customers. To inform and list all known hazards and harms potentially associated with the noted issue below and list risk mitigation factors associated with the use of the product. Hazards Include: Foreign Object and Component Disassociation. Customer actions remained unchanged.
Initiated
2026-02-02
Root cause
Under Investigation by firm
Distribution
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Recall number
Z-1572-2026
Recall status
Open, Classified
CDRH record ID
218642
FDA classified
2026-03-18
Recall event number
98411

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗