Enpath has become aware that some of the Enpath Medical 8 Fr Steerable sheath devices may have non-conforming hemostasis valves in that the valve can leak or the valve was observed to have a small hole.
Source facts
Action
Clinical locations were sent an Urgent Medical Device Recall letter dated 09/02/05 from Enpath Medical, Inc. The letter states the issue, affected lots to be segregated and to discontinue use of. A Bard Electrophysiology Territory Manager will retrieve affected product.
Initiated
2005-08-31
Root cause
Other
Terminated
2006-07-22
Distribution
MA
Recall number
Z-1573-05
Recall status
Terminated
CDRH record ID
41714
FDA classified
2005-09-21
Recall event number
33338