Open original FDA record ↗Recall record
- FDA identifier
- Z-1573-2022
- Company in FDA record
- Mizuho OSI
- FDA record creation date
- Aug 24, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One batch of "regular" femoral hooks were incorrectly etched with the "classic" femoral hook model number.
Source facts
- Action
- On or about 07/05/2022, Muzuho OSI Representative started visiting customers to inform them of a labeling error with the Femur Hook, Right 6850-918 which was incorrectly label with part number 6850-144. Mizuho Representatives were instructed to inform customers: -Over this summer, their facility should have received a Classic Femur Hook, Right 6850-144, either as part of your Hana Table purchase or as a replacement part. But, in error, may have received the other style of femur hook mislabeled with the part number 6850-144. -Hand deliver the Customer Notification Letter and Acknowledge Form. Mizuho Representative are further instructed to: -Inspect the customers' inventory and swap/replace all incorrectly labeled products with the correctly labeled product.
- Initiated
- 2021-07-29
- Root cause
- Process control
- Distribution
- U.S Nationwide Distribution .: CA, FL, GA, MI, NE, NJ, TN, UT, and WA O.U.S.: Canada
- Recall number
- Z-1573-2022
- Recall status
- Open, Classified
- CDRH record ID
- 195027
- FDA record created
- 2022-08-24
- Recall event number
- 90603