In a specific use case, when printing labels for slides, some labels may display wrong information
Source facts
Action
Techidata issued Medical Device Recall Urgent letter via email on 3/27/23. Letter states reason for recall, health risk and action to take:
To prevent any occurrence of the issue, you must not close the current session, or open another one, or perform the printing of another label batch, until the label printing is finished.
Note: printing is finished when the Printing dialog box closes.
software modification was performed to prevent the risks mentioned above.
We kindly ask you to fill the reply form below and return it to us at your earliest convenience.
Initiated
2023-03-27
Root cause
Software design
Distribution
US Nationwide distribution in the state of CA.
Recall number
Z-1573-2023
Recall status
Open, Classified
CDRH record ID
199838
FDA classified
2023-05-10
Recall event number
92184