Open original FDA record ↗Recall record
- FDA identifier
- Z-1573-2026
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Mar 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Source facts
- Action
- On February 18, 2026, the firm notified consignees via Urgent Medical Device Recall letters. Users were instructed to immediately locate, isolate, and stop using the listed affected products. Customers should contact Baxter Technical Support to arrange for the return and correction of their impacted device. Baxter may provide a loaner device upon request while your device is being upgraded and serviced. Distributors should return unopened products in inventory for credit. If affected product was further distributed, the distributor should conduct a consumer-level recall.
- Initiated
- 2026-02-18
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.
- Recall number
- Z-1573-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218746
- FDA classified
- 2026-03-18
- Recall event number
- 98439