Z-1574-2022: Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vant

Recall record Z-1574-2022 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1574-2022
Company in FDA record
Medtronic Neuromodulation
FDA classification date
Aug 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential communication issue in which the Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 may be unable to connect with a Vanta Implantable Neurostimulator (INS) Model 977006 when a patient is implanted with more than one neurostimulator.

Source facts

Action
On about 07/26/2022, Medtronic notified consignees via letter titled "URGENT: MEDICAL DEVICE CORRECTION" sent through courier service. The notification instructed consignees to hold off implanting more than one Vanta INS in a patient until software update is available and while not recommended, if implanting a patient with two (2) Vanta INSs, ensure the INSs are implanted on opposite sides of the body as noted in the Information for Prescribers labeling, and ensure each is configured with the CP App before implanting in a patient. Consignees were asked to complete and return the customer confirmation form provided an contact Medtronic Technical Services at 1-800-707-0933 if troubleshooting does not resolve any issues encountered.
Initiated
2022-07-26
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico. The countries of Austria, Belgium, Canada, Canary Islands, Cayman Islands, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Kingdom.
Recall number
Z-1574-2022
Recall status
Open, Classified
CDRH record ID
195420
FDA classified
2022-08-25
Recall event number
90685

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗