Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.
Source facts
Action
Medtronic issued an URGENT; Medical Device Removal notice on 11/19/2021. The notice explained the issue and the risk and requested the removal of the affected product from use pending return. It also directed the consignee to transfer the notice to other organizations which could be affected.
Initiated
2021-11-19
Root cause
Under Investigation by firm
Distribution
US: IN, CT, NE
Recall number
Z-1574-2023
Recall status
Open, Classified
CDRH record ID
199775
FDA classified
2023-05-10
Recall event number
92142