Skip to content

Public FDA record

Z-1575-2015: NexGen Augment; 5mm Distal Femoral Augment;Part number 00-5990-036-10. Intended to be used in total knee

Recall record Z-1575-2015 for Zimmer, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1575-2015
Company in FDA record
Zimmer, Inc.
FDA classification date
May 1, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The affected lot is being recalled because it was packaged without the locking screw. The locking screw packaged with the augment is used to fix the augment to the femoral component. This has the potential to delay surgery

Source facts

Action
On 4/2/2015, URGENT MEDICAL DEVICE RECALL LOT SPECIFIC notifications were sent to the affected distributors with instructions for locating, quarantining, and returning the affected product. All distributors were notified via electronic mail, and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.
Initiated
2015-04-02
Root cause
Mixed-up of materials/components
Terminated
2015-09-23
Distribution
Distributed in the states of IN, MN, FL, and IL, and the countries of Argentina and Germany.
Recall number
Z-1575-2015
Recall status
Terminated
CDRH record ID
135840
FDA classified
2015-05-01
Recall event number
71009

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗