Skip to content

Public FDA record

Z-1575-2016: Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time,

Recall record Z-1575-2016 for Mindray DS USA, Inc. dba Mindray North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1575-2016
Company in FDA record
Mindray DS USA, Inc. dba Mindray North America
FDA record creation date
Apr 27, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.

Source facts

Action
Mindray DS USA Inc. issued a Medical Device Service Notice dated March 9, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Customers requiring technical assistance were instructed to contact Mindray's Technical Support team at 1-800-288-2121. Mindray will address the systems response to DST in a future software update which will be provided to the customer at no cost.
Initiated
2016-03-09
Root cause
Unknown/Undetermined by firm
Terminated
2018-04-19
Distribution
Nationwide Distribution
Recall number
Z-1575-2016
Recall status
Terminated
CDRH record ID
144763
FDA record created
2016-04-27
Recall event number
73693

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗