Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.
Source facts
Action
Mindray DS USA Inc. issued a Medical Device Service Notice dated March 9, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Customers requiring technical assistance were instructed to contact Mindray's Technical Support team at 1-800-288-2121. Mindray will address the systems response to DST in a future software update which will be provided to the customer at no cost.
Initiated
2016-03-09
Root cause
Unknown/Undetermined by firm
Terminated
2018-04-19
Distribution
Nationwide Distribution
Recall number
Z-1575-2016
Recall status
Terminated
CDRH record ID
144763
FDA record created
2016-04-27
Recall event number
73693