Open original FDA record ↗Recall record
- FDA identifier
- Z-1576-2016
- Company in FDA record
- Jas Diagnostics/Drew Scientific
- FDA record creation date
- Apr 27, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The reagent pack may result in elevated platelet background during Drew3 Hematology Analyzer routine check.
Source facts
- Action
- JAS Diagnostics/ Drew Scientifics sent an Important Product Advisory letter dated March 10, 2016, to all affected customers. The domestic customers were notified via letter and international customers were notified via email. Customers were instructed to contact Drew Scientific Inc. Customer Service or Technical Service at 1-800-433-0945 for product replacement. The firm suggested a temporary workaround of performing five background runs three times per day in order to ensure background counts are running within specification.
- Initiated
- 2016-03-10
- Root cause
- Process control
- Terminated
- 2019-08-21
- Distribution
- Worldwide Distribution - US Distribution including the states of : AL, AR, CA, FL, GA, IA, IN, KS, KY, LA, MA, MN, MO, MS, NC, NE, NY, SC, TN, and TX. and to the countries of : Bolivia, Cairo, Columbia, Ecuador, Italy, Moscow, Netherlands, Pakistan, Peru, Slovenia, Spain, and Venezuela.
- Recall number
- Z-1576-2016
- Recall status
- Terminated
- CDRH record ID
- 144664
- FDA record created
- 2016-04-27
- Recall event number
- 73650