Open original FDA record ↗Recall record
- FDA identifier
- Z-1576-2023
- Company in FDA record
- Diagnostica Stago, Inc.
- FDA classification date
- May 10, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).
Source facts
- Action
- Customers were contacted via telephone and an URGENT: MEDICAL DEVICE CORRECTION notification letter beginning 3/22/23. Actions: Stago downgraded your firmware to the previous firmware version as a precaution. Your downgrade has already occurred. Consequently, the risk of this bug only existed during the time period shown above (i.e. from new firmware installation date to the downgrade date). As patient results are interpreted in association with other biological tests, we leave the decision to review previous patient results based on the complete clinical context to your discretion. Our Customer Service team is available to help you carry out your impact analysis. For additional information, please contact your Stago Hotline at 1-800-725-0607.
- Initiated
- 2023-03-22
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.
- Recall number
- Z-1576-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199408
- FDA classified
- 2023-05-10
- Recall event number
- 92007