- Recall classification
- Class II
Reason stated in the source
When Epiq 7 Ultrasound System, WITH the Pediatric Cardiology option, is set up to Metric and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patients Body Surface Area (BSA).
Source facts
- Action
- Philips sent the Urgent-Medical Device Correction-Philips Model EPIQ5 and Ultrasound system letter, dated 2015 APRIL 10, and the Urgent-Medical Device Correction-Philips Model EPIQ7 and Ultrasound system letter, dated 2015 APRIL 10, letters to their consignees. These systems WITH Pediatric Cardiology option are affected. Philips issued two Field Change Orders (FCO) to their Field Service Engineer to install upgraded software: -FCO 79500336 for EPIQ5 -FCO 79500337 for EPIQ7 Customers with questions can call Philips Customer Service at 1-800-722-9377.
- Initiated
- 2015-04-10
- Root cause
- Software design
- Terminated
- 2016-02-29
- Distribution
- Worldwide Distribution - US (nationwide and DC) and Internationally to Norway, India, New Zealand, Bahrain, Sweden, Italy, Australia, France, Netherlands, Switzerland, Germany, Austria, and Turkey.
- Recall number
- Z-1579-2015
- Recall status
- Terminated
- CDRH record ID
- 136056
- FDA classified
- 2015-05-05
- Recall event number
- 71084