Open original FDA record ↗Recall record
- FDA identifier
- Z-1579-2023
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- May 11, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Automated dispensing cabinet software is experiencing: 1) ES device download failure resulting in partial patient information loss that leads to 2) partial loss of patient transaction data, seen as a server upload processing error. Outdated/incorrect information display may contribute to the removal of medication to which the patient is allergic, incorrect dose amount, or discontinued medication.
Source facts
- Action
- On 4/5/23, correction notices were distributed to Biomedical Engineering, Risk Management, Nursing, Pharmacy Directors who were informed of the following: 1) Remind users that automated dispensing cabinet devices are not a replacement for the medical record (e.g., medication orders, medication administration record (MAR), and patient allergy information) and are not to be relied on as the source of truth for medication administration. Consult the hospital medical record before removing any medication. 2. Please ensure this letter is forwarded to applicable facilities if devices have been transferred and that all appropriate personnel have been made aware of this issue. 3. Complete and return the response form to BDRC24@bd.com Recalling Firm Actions: 1. For customers who have reported an upload processing failure, the support team can provide a copy of affected data that is unable to be moved to the ES Server data repository. 2. A software update is available. The recalling firm will contact you to schedule a time to install the software update. For further assistance contact Customer Support at 1-866-583-8783, 5:00am PST to 4:00pm PST, Monday - Friday; or Technical Support, bd.com/self-service, 1-800-727-6102, 24 hours, 7 days a week
- Initiated
- 2023-04-05
- Root cause
- Software Design Change
- Distribution
- US Nationwide distribution in the states of KS, NY, NC, LA, MA, MI.
- Recall number
- Z-1579-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199540
- FDA classified
- 2023-05-11
- Recall event number
- 92061