Open original FDA record ↗Recall record
- FDA identifier
- Z-1580-2015
- Company in FDA record
- Viewray Incorporated
- FDA classification date
- May 5, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The software was not correctly using the RT (Radiation Therapy) to MR (Magnetic Resonance image) coordinate correction for non HFS (Head First Supine) patient orientations, resulting in slice mismatch error.
Source facts
- Action
- On January 15, 2014, ViewRay personnel discussed the investigation of the issue (slice mismatch error) with the customer. The firm advised the customer that until the issue was corrected, to not treat patients in non HFS orientations, or to use the MR acquired from the ViewRay system as the primary plan series for non HFS scans. ViewRay installed software version 3.4.5.28 on January 20 and 21, 2014, in order to prevent this slice mismatch error from occurring in any patient position.
- Initiated
- 2015-01-15
- Root cause
- Software Design Change
- Terminated
- 2015-09-23
- Distribution
- US distribution to MO.
- Recall number
- Z-1580-2015
- Recall status
- Terminated
- CDRH record ID
- 135930
- FDA classified
- 2015-05-05
- Recall event number
- 71050