Skip to content

Public FDA record

Z-1580-2020: LN 3R70-01; software version 2.6.2 and earlier. The software is used with the Alinity i (LN 03R65-01) pr

Recall record Z-1580-2020 for Abbott Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1580-2020
Company in FDA record
Abbott Gmbh & Co. KG
FDA record creation date
Mar 26, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abbott has identified potential performance issues for the Alinity ci -series Software version 2.6.2 and earlier.

Source facts

Action
The firm initiated the recall letter on January 24, 2020. The "Product Correction, Urgent - Immediate Action Required" notice identified the correction of the Alinity ci-series Software version 2.6.2 and earlier due to potential performance issues. the notice informed that Abbott is releasing Alinity ci-series Software version 3.1.0 to correct the issues. The notice provided the necessary actions required to be performed until the mandatory upgrade to SW 3.1.0 is complete. The notice referenced that details are included in table1 below. the notice detail that an Abbott representative will schedule a mandatory upgrade of your Alinity ci-series to software version 3.1.0. The notice also informed the customer if you have forwarded the product listed above to other laboratories, please inform them of the Product Correction and provide them a copy of this letter. The notice further instructs the customer to please retain this letter for your laboratory records The notice instructed the customers to please Abbott's Customer Service at 1 877 4ABBOTT(available 24 hours a day, 7 days a week). For customers outside the U.S., to please contact your local area Customer Service.
Initiated
2020-01-24
Root cause
Software design
Terminated
2023-08-08
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, LA, MA, MD, ME, MI, MN, MC, MS, NC, NJ, NY, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and countries of Albania, Argentina, Australia, Austria, Bahrain, Bangledesh, Belgium, Bermuda, Bosnia, Botswana, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Madagascar, Malaysia, Mexico, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago Turkey, UEA, UK, Ukraine, Uruguay, Vietnam.
Recall number
Z-1580-2020
Recall status
Terminated
CDRH record ID
179503
FDA record created
2020-03-26
Recall event number
84831

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗