Z-1580-2023: DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with

Recall record Z-1580-2023 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1580-2023
Company in FDA record
Smiths Medical ASD Inc.
FDA classification date
May 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient.

Source facts

Action
Smiths Medical issued an URGENT MEDICAL DEVICE RECALL notice on 04/05/2023 by mail (traceable method). The notice explained the issue, potential risk, and request the following actions be taken: "Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. If you have expired product, do not use. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with this letter. Contact Sedgwick at 1-844-861-6220 (MF, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product." "If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Sedgwick at 1-844-861-6220 (M-F, 8am-5pm ET) to obtain a response form."
Initiated
2023-04-05
Root cause
Labeling False and Misleading
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, New Zealand, and Hong Kong.
Recall number
Z-1580-2023
Recall status
Open, Classified
CDRH record ID
199337
FDA classified
2023-05-12
Recall event number
91975

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗