Cribs sold in the USA are missing two access door warning labels.
Source facts
Action
On February 18. 2026, the firm notified customers through IMPORTANT MEDICAL DEVICE CORRECTION letters. Customers were notified of the missing warning labels, which read,
"WARNING: Never leave a child unsupervised when the movable slide is open or not secured
Refer to Instructions for Use for product care, maintenance, and cleaning."
Customers were instructed to locate the units listed on the attached business reply form and identify the address where they can be serviced, and return the completed form to Stryker. Stryker will then contact consignees to arrange for the application of the missing label.
In the interim, until the service can be performed, all pertinent warnings and
safety instructions may be found in the product manual: https://techweb.stryker.com/Stretcher/FL19/1607/Operations/1900-109-
001AC.0.pdf
Initiated
2026-02-18
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-1580-2026
Recall status
Open, Classified
CDRH record ID
218684
FDA classified
2026-03-19
Recall event number
98426