Open original FDA record ↗Recall record
- FDA identifier
- Z-1581-2026
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Mar 19, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to adapter not being properly aligned or locked in place.
Source facts
- Action
- On 02/18/2026 and 02/24/2026, the firm sent via email an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" letter to customers informing them that Elekta has become aware that a Quality Assurance (QA) procedure was initiated while the adapter was not properly locked to the Patient Positioning System of the Leksell Gamma Knife and as a result, the QA check failed. Customers are instructed to: To prevent the risk of incorrect locking, it is important to always check that the patient is properly attached to the Leksell Gamma Knife. The adapter shall be perfectly aligned with the docking device. Reference instructions for use and warnings: Warning 4.1 - After locking the docking device, make sure that the adapter is correctly attached and securely locked to the patient positioning system. if the adapter is not correctly locked, it may affect the accuracy and lead to clinical mistreatment, or cause patient injury. All Customer must print/save this page and add it to their existing IFU file. For questions - Contact Region America at Application.Support.NA@elekta.com or telephone +1 855-693-5358
- Initiated
- 2026-02-18
- Root cause
- Use error
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
- Recall number
- Z-1581-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218768
- FDA classified
- 2026-03-19
- Recall event number
- 98443