Z-1581-2026: Brand Name: Leksell Gamma Knife Product Description: Leksell Gamma Knife Perfexion Leksell Gamma Knife Ic

Recall record Z-1581-2026 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1581-2026
Company in FDA record
Elekta, Inc.
FDA classification date
Mar 19, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to adapter not being properly aligned or locked in place.

Source facts

Action
On 02/18/2026 and 02/24/2026, the firm sent via email an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" letter to customers informing them that Elekta has become aware that a Quality Assurance (QA) procedure was initiated while the adapter was not properly locked to the Patient Positioning System of the Leksell Gamma Knife and as a result, the QA check failed. Customers are instructed to: To prevent the risk of incorrect locking, it is important to always check that the patient is properly attached to the Leksell Gamma Knife. The adapter shall be perfectly aligned with the docking device. Reference instructions for use and warnings: Warning 4.1 - After locking the docking device, make sure that the adapter is correctly attached and securely locked to the patient positioning system. if the adapter is not correctly locked, it may affect the accuracy and lead to clinical mistreatment, or cause patient injury. All Customer must print/save this page and add it to their existing IFU file. For questions - Contact Region America at Application.Support.NA@elekta.com or telephone +1 855-693-5358
Initiated
2026-02-18
Root cause
Use error
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
Recall number
Z-1581-2026
Recall status
Open, Classified
CDRH record ID
218768
FDA classified
2026-03-19
Recall event number
98443

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗