Cuff may underinflate, or device may stop working during use. Cardioassist software version 4.3.1 utilized a signal from the plethysmograph that was found not to be sensitive enough due to the shape of the waveform in certain circumstances where there was excessive motion artifact.
Source facts
Action
Service visit to customer in May 2009 to upgrade and validate CardioAssist 4000 software to version 4.3.5.
Initiated
2009-04-09
Root cause
Software design
Terminated
2012-08-27
Distribution
Florida
Recall number
Z-1582-2009
Recall status
Terminated
CDRH record ID
82561
FDA classified
2009-07-07
Recall event number
52223