Skip to content

Public FDA record

Z-1584-2025: Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog N

Recall record Z-1584-2025 for Channel Medsystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1584-2025
Company in FDA record
Channel Medsystems, Inc.
FDA classification date
Apr 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device

Source facts

Action
On 03/20/2025, the firm emailed customer an "URGENT MEDICAL DEVICE CORRECTION" Letter informing them that the evaporator gasket inside the handle of the device may not seal properly, which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device. Customers are instructed to: 1. Check your internal inventory to determine if you have any affected devices. 2. Stop use of affected devices. 3. Vent the affected devices by following the attached instructions and then discard the devices according to your facility s procedures. 4. Complete and return the customer acknowledgement form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore, please complete the form even if you no longer have any of the affected devices in your physical inventory. 5. Circulate this letter to all other relevant parties, as needed. For questions, contact Channel Medsystems at safetynotice@cerene.com or at (510)338-9301 Monday thru Friday 8:00 AM to 5:00 PM Pacific Time
Initiated
2025-03-20
Root cause
Under Investigation by firm
Distribution
U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX.
Recall number
Z-1584-2025
Recall status
Open, Classified
CDRH record ID
213018
FDA classified
2025-04-16
Recall event number
96485

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗