Angiomed Material Review Board reviewing the manufacturing process observed an increase in the internal scrap rate (failure mode: broken stent struts) of nitinol stents and determined the root cause to be a manufacturing change from polymer to aluminum fixtures.
Source facts
Action
Recall notification letter dated July 21, 2005, was sent to Hospital and Physician consignees.
Initiated
2005-07-21
Root cause
Other
Terminated
2008-10-23
Distribution
Nationwide
Recall number
Z-1585-05
Recall status
Terminated
CDRH record ID
41079
FDA classified
2005-09-24
Recall event number
32921