Skip to content

Public FDA record

Z-1585-2016: X-Ray Diagnostic Table, Model No. YSF-300 Use in combination with an X-ray high voltage generator, an X-r

Recall record Z-1585-2016 for Shimadzu Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1585-2016
Company in FDA record
Shimadzu Medical Systems
FDA classification date
Apr 28, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Report of unintentional movement of table to reverse tilting.

Source facts

Action
Shimadzu sent an Urgent Voluntary Medical Device Recall Notice to all affected customers on April 8, 2016. Shimadzu Corporation has identified a potential issue in the Shimadzu X-ray Diagnostic Table. The letter informs the customers of the problems identified and the actions to be taken. Customers with questions or concerns are instructed to contacted Shimadzu National Technical Support at (800) 228-1429. Shimadzu will send out an updated customer notification letter to inform customers that the liquid penetration may cause an electrical short circuit, causing the table to unintentionally tilt in reverse. If this unintentional reverse tilting occurs, there is a potential risk of injury to the patient and/or the user. For questions regarding this recall call 310-217-8855, ext 139.
Initiated
2016-03-31
Root cause
Device Design
Terminated
2017-01-02
Distribution
Nationwide Distribution
Recall number
Z-1585-2016
Recall status
Terminated
CDRH record ID
144860
FDA classified
2016-04-28
Recall event number
73745

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗