Open original FDA record ↗Recall record
- FDA identifier
- Z-1585-2025
- Company in FDA record
- Conavi Medical Inc.
- FDA classification date
- Apr 23, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Due to manufacturing issues there is a potential for the catheter sheath to detach.
Source facts
- Action
- On 03/12/2025, the firm sent via email an "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" Letter informing customer that the firm has become aware that the sheath of the Novasight Hybrid catheter detached and left in the patient's body during a clinical case but was retrieved during the procedure. Customers are instructed to: Customers shall stop/cease use of the affected the Novasight Hybrid catheters and remove them from service and return immediately to Conavi Medical Inc. This applies to all Novasight Hybrid catheters shipped to U.S.A. regardless of expiry dates. Acknowledge receipt of this recall notification by completing and returning the "Acknowledgement" section of this notification by email to: julie@conavi.com. For questions or assistance - Contact information: Director of Product Management and Marketing Telephone # (416)-483-0100 Ext 109, Email: julie@conavi.com Days/Hours Available for calls: Monday through Friday, 9:00 AM to 4:30 PM Eastern Time.
- Initiated
- 2025-03-12
- Root cause
- Process control
- Distribution
- US distribution to states of: GA, NY, NJ, OH; and OUS (Foreign) to countries of: Canada and China
- Recall number
- Z-1585-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212885
- FDA classified
- 2025-04-23
- Recall event number
- 96444