Skip to content

Public FDA record

Z-1585-2025: Brand Name: Novasight Hybrid System Product Name: Novasight Hybrid Catheter Model/Catalog Number: TA-06-0

Recall record Z-1585-2025 for Conavi Medical Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1585-2025
Company in FDA record
Conavi Medical Inc.
FDA classification date
Apr 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to manufacturing issues there is a potential for the catheter sheath to detach.

Source facts

Action
On 03/12/2025, the firm sent via email an "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" Letter informing customer that the firm has become aware that the sheath of the Novasight Hybrid catheter detached and left in the patient's body during a clinical case but was retrieved during the procedure. Customers are instructed to: Customers shall stop/cease use of the affected the Novasight Hybrid catheters and remove them from service and return immediately to Conavi Medical Inc. This applies to all Novasight Hybrid catheters shipped to U.S.A. regardless of expiry dates. Acknowledge receipt of this recall notification by completing and returning the "Acknowledgement" section of this notification by email to: julie@conavi.com. For questions or assistance - Contact information: Director of Product Management and Marketing Telephone # (416)-483-0100 Ext 109, Email: julie@conavi.com Days/Hours Available for calls: Monday through Friday, 9:00 AM to 4:30 PM Eastern Time.
Initiated
2025-03-12
Root cause
Process control
Distribution
US distribution to states of: GA, NY, NJ, OH; and OUS (Foreign) to countries of: Canada and China
Recall number
Z-1585-2025
Recall status
Open, Classified
CDRH record ID
212885
FDA classified
2025-04-23
Recall event number
96444

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗