Open original FDA record ↗Recall record
- FDA identifier
- Z-1586-2026
- Company in FDA record
- Hologic, Inc.
- FDA classification date
- Mar 19, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to product exhibiting potential to generate either invalid or false negative results.
Source facts
- Action
- On 02/18/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter to customers informing them that Hologic has identified that the Aptima HPV Assay kits from ML 929785 exhibit the potential to generate either invalid results for samples and calibrators or false negative results for low analyte samples when reagents are not used on the day of reconstitution. Customers are instructed: 1. Discontinue use of the Aptima HPV Assay kits from ML 929785. 2. Segregate and count the number of affected product on-hand and dispose according t the local guidelines and regulations. 3.Fill out the Customer Response Form using the link https://iqvia-response.my.site.com/mt/fca?cid=hpv-929785 4. Hologic will process and ship the replacement kits For Questions or concerns, contact Hologic Technical Support at +1 888-484-4747 or +1 858-410-8511 during hours 5 am to 5 pm PT or by email at molecularsupport@hologic.com
- Initiated
- 2026-02-18
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Nationwide distribution in the states of CA, FL. IA, IL, MO, NC, NM, PR, and TN.
- Recall number
- Z-1586-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218748
- FDA classified
- 2026-03-19
- Recall event number
- 98441