- Action
- Field Safety Notices were distributed on 3/14/18. The notices instructed customers:
To prevent occurrence of the issue customers shall restart the system at least once a day.
Customer shall ensure that all staff with access to the affected systems are informed of the contents of this Field Safety Notice.
A copy of this Field Safety Notice shall be placed together with the documentation of the system until the system has been corrected by Philips.
The problems will be resolved by a software update, which is expected to be available by the second half of 2018.
You will be notified by your local Philips representative when the software update is available for installation.
If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377, 8am to 8pm (EST) Monday thru Friday.
- Initiated
- 2018-03-14
- Root cause
- Software design
- Terminated
- 2024-04-29
- Distribution
- Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
The product was distributed to the following countries: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Croatia, Cuba, Czech Republic, Denmark, Estonia, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Republic of Korea, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, Norway, Pakistan, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, and United Kingdom.
- Recall number
- Z-1589-2018
- Recall status
- Terminated
- CDRH record ID
- 163417
- FDA record created
- 2018-05-01
- Recall event number
- 79770