Z-1591-2026: Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impel

Recall record Z-1591-2026 for Abiomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1591-2026
Company in FDA record
Abiomed, Inc.
FDA classification date
Mar 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

Source facts

Action
Abiomed notified international consignees on about 02/16/2026 via emailed CRITICAL PRODUCT CORRECTION letter. Consignees were instructed to cease use of the Automated Impella Controller (AIC) with an Impella 5.5 S1 (product code 0550-0007) unless AIC software is updated to v10.1 or above, identify current software in the lower left corner of the initial AIC display screen, contact your local Abiomed representative to schedule your software update, review and complete the customer response form provided and post this notification in a visible area for awareness. Consignees were also instructed to forward this notification to anyone within the facility that should be aware and forward the notification to consignees if further distributed.
Initiated
2026-02-16
Root cause
Software design
Distribution
International distribution in the country of Australia, Canada, France, Germany, Kuwait.
Recall number
Z-1591-2026
Recall status
Open, Classified
CDRH record ID
218584
FDA classified
2026-03-27
Recall event number
98393

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗