Skip to content

Public FDA record

Z-1592-2008: InCompass Spinal Fixation System T-Handle Driver; Part #2151-4 (9.5" shaft) and 2151-5 (13" shaft); Non-s

Recall record Z-1592-2008 for Abbott Spine. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1592-2008
Company in FDA record
Abbott Spine
FDA classification date
Sep 1, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Reports received regarding disassembly of T-Handle Driver components during surgery dropping the closure top into the surgical site.

Source facts

Action
Firm sent notification letters to sales agents via letter on 03/24/08. Enclosed with the notification letter, the firm provided updated product inserts for the InCompass and PathFinder Spinal Fixation System (Cat. #2199-0002-LB Rev. H and #1199-0002-LB Rev. H, respectively), as well as the "Disassembly and Assembly Instructions for PathFinder & InCompass T-Handle Drivers" (Cat #1199-0040-MKC Rev. A). Sales Agents instructed to distribute updated inserts and instructions to all locations that handle the drivers for cleaning and to destroy all copies of old inserts. If you have questions or have not received the recall notice from the firm, contact Kirstin Widding at 512-533-1852.
Initiated
2008-03-24
Root cause
Error in labeling
Terminated
2009-12-02
Distribution
Nationwide Distribution --- including states of AR, AZ, CA, CT, CO, FL, IL, LA, MD, MI, MO, MN, MT, NC, NM, NJ, NY, OH, OK, OR, PA, SD, TX, VA and WI.
Recall number
Z-1592-2008
Recall status
Terminated
CDRH record ID
69480
FDA classified
2008-09-01
Recall event number
47424

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗