Open original FDA record ↗Recall record
- FDA identifier
- Z-1592-2024
- Company in FDA record
- ACTIM OY
- FDA classification date
- Apr 17, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.
Source facts
- Action
- Actim Oy issued a MEDICAL DEVICE RECALL notice to its sole distributor on 11/23/2023 via email. The notice explained the issue, potential risk, and requested that users refrain from using surgical lubricants when collecting the vaginal specimen and that vaginal specimens do not contain significant amounts of personal lubricant from patient's recent personal use. The instructions for use will be updated according to the latest test results of surgical and personal lubricant interference. In the interim, a separate note will be included in the test kit regarding the correction of the interference claims and guidance of lubricant use.
- Initiated
- 2023-11-23
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the state of Connecticut.
- Recall number
- Z-1592-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205627
- FDA classified
- 2024-04-17
- Recall event number
- 93845