Open original FDA record ↗Recall record
- FDA identifier
- Z-1593-2008
- Company in FDA record
- Abbott Spine
- FDA classification date
- Sep 1, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Reports received regarding disassembly of T-Handle Driver components during surgery dropping the closure top into the surgical site.
Source facts
- Action
- Firm sent notification letters to sales agents via letter on 03/24/08. Enclosed with the notification letter, the firm provided updated product inserts for the InCompass and PathFinder Spinal Fixation System (Cat. #2199-0002-LB Rev. H and #1199-0002-LB Rev. H, respectively), as well as the "Disassembly and Assembly Instructions for PathFinder & InCompass T-Handle Drivers" (Cat #1199-0040-MKC Rev. A). Sales Agents instructed to distribute updated inserts and instructions to all locations that handle the drivers for cleaning and to destroy all copies of old inserts. If you have questions or have not received the recall notice from the firm, contact Kirstin Widding at 512-533-1852.
- Initiated
- 2008-03-24
- Root cause
- Error in labeling
- Terminated
- 2009-12-02
- Distribution
- Nationwide Distribution --- including states of AR, AZ, CA, CT, CO, FL, IL, LA, MD, MI, MO, MN, MT, NC, NM, NJ, NY, OH, OK, OR, PA, SD, TX, VA and WI.
- Recall number
- Z-1593-2008
- Recall status
- Terminated
- CDRH record ID
- 69481
- FDA classified
- 2008-09-01
- Recall event number
- 47424