Open original FDA record ↗Recall record
- FDA identifier
- Z-1594-2021
- Company in FDA record
- WillowWood Global Llc
- FDA record creation date
- May 14, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
Source facts
- Action
- On 04/20/2021, Willow Wood issued an Urgent Medical Device Removal notice to customer via letter notifying them it was discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
- Initiated
- 2021-04-20
- Root cause
- Process change control
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, KY, LA, MA, ME, MN, MO, MS, NJ, NM, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and the countries of Germany, Netherlands
- Recall number
- Z-1594-2021
- Recall status
- Open, Classified
- CDRH record ID
- 187128
- FDA record created
- 2021-05-14
- Recall event number
- 87825