The firm is conducting a sub recall of kits containing ConMed Goldline Rocker Switch Pencils. The pencils activation switch may remain in the "on" position after the switch has been depressed with excessive force, thereby continuously activating the pencil.
Source facts
Action
Centurion sent out an "Urgent Product Recall Notice" to one customer on 4/9/2010. The customer was instructed to discontinue use of the product. The customer was also instructed to return any instock product .
Initiated
2010-04-09
Root cause
Nonconforming Material/Component
Terminated
2010-10-25
Distribution
TX
Recall number
Z-1595-2010
Recall status
Terminated
CDRH record ID
90519
FDA classified
2010-05-12
Recall event number
55305
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.