Open original FDA record ↗Recall record
- FDA identifier
- Z-1595-2011
- Company in FDA record
- AGA Medical Corporation
- FDA classification date
- Mar 8, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
AGA Medical is initiating a voluntary recall of a specific lot of mislabeled AMPLATZER Sizing Balloon II. AGA Medical received 2 reports from customers who received sizing balloons that were labeled as 34 mm; however, the package contained 24 mm devices. The proper sizing (24 mm) was printed on the device hub visible through the product package.
Source facts
- Action
- AGA Medical Corporation sent an Urgent Medical Device Voluntary Recall letter, dated December 17, 2010, via e-mail to all affected customers. A hard copy was sent via Fed Ex to customers on December 20, 2010. The letter identified the product, described the issue, and the action that should be taken by the customer. Customers were requested to cease distribution and use of the devices, check inventory for unused devices, and complete and return a response form to facilitate return of the affected devices to AGA. For questions regarding this recall call 1-763-531-3175.
- Initiated
- 2010-12-17
- Root cause
- Employee error
- Terminated
- 2012-03-29
- Distribution
- Worldwide Distribution - USA including AZ, CA, KY, PA, SC, TX and OUS to include: Argentina, Germany, Greece, Netherlands, Peru, Poland, Portugal, Slovakia, United Kingdom
- Recall number
- Z-1595-2011
- Recall status
- Terminated
- CDRH record ID
- 97502
- FDA classified
- 2011-03-08
- Recall event number
- 57819