Z-1595-2023: Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062

Recall record Z-1595-2023 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1595-2023
Company in FDA record
Steris Corporation
FDA classification date
May 17, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

Source facts

Action
An URGENT MEDICAL DEVICE FIELD CORRECTION dated 3/28/23 was sent to customers. STERIS Action: All affected Customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their unit(s) to perform the correction. User Action: Users can continue to use their Reliance Vision Multi-Chamber Washer/Disinfector prior to completion of the correction. Should your unit alarm Drying Temperature Too High, abort the cycle, turn off the power to the unit, and contact STERIS. The competent (regulatory) authority of your country has been informed of this notice. Customer response is not required. Users are reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact Senior Product Manager at 440-392-7571, STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.
Initiated
2023-04-10
Root cause
Nonconforming Material/Component
Distribution
Worldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of: China, Lithuania, Saudi Arabia, Korea, Republic of, Canada, Taiwan, United Kingdom, Spain, Australia, New Zealand, Spain, Brazil, Bahrain, Switzerland, Great Britain, South Africa, Hong Kong, Qatar.
Recall number
Z-1595-2023
Recall status
Open, Classified
CDRH record ID
199457
FDA classified
2023-05-17
Recall event number
92044

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗