Reason stated in the source
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Source facts
- Action
- Sizewise Manufacturing issued an URGENT Voluntary Medical Device Recall to its consignees on 01/29/2024 by letter. The notice explained the issue, potential risk, and requested the discontinuation of use and disposal of the device. The firm sent a follow-up notice on 03/07/2024 asking the distributors to further notify their customers (users of the device).
- Initiated
- 2024-01-25
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, UK.