LifeScan brand OneTouch Ultra Meter - Owner's Booklets are missing instructions on how to set the Units of Measurement (UOM)--mg/dL versus mmol/L. This error has the potential to lead to acute hyperglycemic complications if the user fails to realize that 18mg/dL of glucose is equal to 1 (one) mmol/L.
Source facts
Action
Firm contacted all accounts by letters of 11/21/03, 2/4/2004, 2/17/04 and 2/9-10/2004 providing important information regarding the owner''s booklet.
Initiated
2003-11-21
Root cause
Other
Terminated
2005-11-29
Distribution
The recall product was distributed in the U.S. only.
LifeScan manufactured and distributed OneTouch Ultra System Kits with the affected OneTouch Ultra Meter Owner''s Booklets between March 31, 2003 and October 3, 2003.
Recall number
Z-1596-05
Recall status
Terminated
CDRH record ID
38552
FDA classified
2005-09-29
Recall event number
31815
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.