Skip to content

Public FDA record

Z-1596-2010: ARCHITECT HAVAB-M Reagent A chemiluminescent microparticle immunoassay (CMIA) for the qualitative detect

Recall record Z-1596-2010 for Abbott Diagnostics International, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1596-2010
Company in FDA record
Abbott Diagnostics International, Ltd.
FDA classification date
May 19, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a potential to generate a false grayzone or reactive HAVAB-M result when the ARCHITECT AUSAB assay precedes the ARCHITECT HAVAB-M assay.

Source facts

Action
Direct accounts (hospitals, private laboratories, and other institutions) were sent a Product Correction letter, dated 12/30/2009. The letter identified the affected product, explained the issue and patient impact, and listed several necessary actions. Customers were instructed to follow one of 4 different alternatives to protect the integrity of test results. The letter is to be retained for laboratory records and a copy should be sent to other laboratories that kits were forwarded to. Questions regarding this information should be directed to Customer Service at 1-877-4ABBOTT.
Initiated
2009-12-30
Root cause
Pending
Terminated
2011-08-31
Distribution
Nationwide Distribution -- Including Puerto Rico.
Recall number
Z-1596-2010
Recall status
Terminated
CDRH record ID
90201
FDA classified
2010-05-19
Recall event number
55165

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗