Z-1596-2016: VITEK 2 Gram Positive Susceptibility Test Cards (AST-P605), REF 22 325. 20 Cards/Kit. bioMerieux, Inc, Du

Recall record Z-1596-2016 for Biomerieux Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1596-2016
Company in FDA record
Biomerieux Inc
FDA classification date
May 3, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline.

Source facts

Action
Biomerieux, Inc. sent an Urgent Notice Product Correction on February 24, 2016, to all affected customers. The recalling firm notified their foreign subsidiary of the Field Safety Corrective Action (FSCA) by email. The local subsidiary in the country impacted has informed users impacted and provided instructions to resolve the issue. Customers were instructed to confirm the Field Safety Corrective Action letter has been distributed and reviewed by all appropriate personnel within their organization. For the referenced test kit (VITEK 2 ASTP605, Lot 485368220), refrain from reporting Minocycline (MNO) results for any organism until a new lot is implemented in your laboratory. o Please note that a bioART rule can be created to suppress the analysis/reporting of Minocycline (reference VITEK 2 Online Software User Manual). The bioART rule can be disabled upon implementation of a new ASTP605 lot number. Customers were instructed tostore the letter with their bioMrieux instrument documentation. Complete the attached Acknowledgement Form and return it to their local bioMrieux representative. For questions regarding this recall call 314-731-8526.
Initiated
2016-02-24
Root cause
Other
Terminated
2017-05-26
Distribution
Worldwide Distribution to China Taiwan region; No US/military/government distribution.
Recall number
Z-1596-2016
Recall status
Terminated
CDRH record ID
144536
FDA classified
2016-05-03
Recall event number
73585

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗