Open original FDA record ↗Recall record
- FDA identifier
- Z-1597-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- May 2, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view.
Source facts
- Action
- Customers were notified via letter on approximately 01/26/2018. Instructions included to ensure proper positioning of the PET QC Phantom when performing the PET Quality Control Procedure by following section 6.3.1 in the Biograph Horizon Operator's Manual for Examination and Acquisition, complete the software update to PETsyngo VJ20B, ensure the recall notice is placed in the Biograph Horizon Operator's Manual for Examination and Acquisition and disseminated to all operators of the Biograph Horizon. If the equipment has been sold to another customer, they are instructed to provide the recall notice to the new owner.
- Initiated
- 2018-01-26
- Root cause
- Software Design Change
- Terminated
- 2018-10-12
- Distribution
- Worldwide distribution. US nationwide, Australia, France, India, Japan, and Netherlands.
- Recall number
- Z-1597-2018
- Recall status
- Terminated
- CDRH record ID
- 164139
- FDA record created
- 2018-05-02
- Recall event number
- 79988