Open original FDA record ↗Recall record
- FDA identifier
- Z-1599-2020
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Feb 27, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The assay may exhibit a positive bias in QC and patient results within the 10-day pack calibration interval.
Source facts
- Action
- On February 11, 2020, the firm issued an Urgent Medical Device Correction to all US customers. Customers who experience the issue were instructed to perform a pack calibration every 2 days or more frequently based on the laboratory's Quality Control performance. The firm is working to address the issue in a future software version. Customers are requested to complete and return the effectiveness check questionnaire. If you have any questions, contact your local Siemens Healthineers technical support representative.
- Initiated
- 2020-02-14
- Root cause
- Under Investigation by firm
- Terminated
- 2024-02-22
- Distribution
- Domestic distribution nationwide. Foreign distribution to Australia, Austria, Bahamas, Belgium, Canada, Curacao, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kenya, Latvia, Mexico, Netherlands, Norway, Poland, Portugal, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., Uruguay, United Kingdom, and Vietnam.
- Recall number
- Z-1599-2020
- Recall status
- Terminated
- CDRH record ID
- 179926
- FDA classified
- 2020-02-27
- Recall event number
- 85020