If used, the optional Swap feature copies programming parameters from the external pulse generator (EPG) as well as unintended calibration data to the implantable pulse generator (lPG). This may result in the user being unable to recharge the lPG.
This safety notice only applies to CPs Model 4500.
Source facts
Action
Nuvectra sent an "Urgent Field Safety Notice" letter dated April 6, 2016 to customers. The letter identified the affected product, problem, recommendations and Mitigations, and actions to be taken.
Initiated
2016-04-06
Root cause
Software design
Terminated
2017-09-29
Distribution
Germany
Recall number
Z-1600-2016
Recall status
Terminated
CDRH record ID
145307
FDA classified
2016-05-03
Recall event number
73884