Skip to content

Public FDA record

Z-1600-2023: Fisher Wallace Stimulator Models FW-100 and FW-200

Recall record Z-1600-2023 for Fisher Wallace Laboratories Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1600-2023
Company in FDA record
Fisher Wallace Laboratories Inc.
FDA classification date
May 17, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the batteries are installed incorrectly, the product may overheat, possibly leading to pain, discomfort, irritation, or even burns during use of the device.

Source facts

Action
An URGENT MEDICAL DEVICE CORRECTION dated 4/6/23 was sent to customers. Short term actions to be taken by the User: 1. Always ensure the batteries are inserted correctly into the device (see above diagram). 2. Do not leave batteries in the device when the device is not in use (e.g. in between a morning and evening treatment). 3. Do not fall asleep while using the device because corrective actions cannot be taken when the user is asleep. 4. If the device exhibits excess heat, burning smell, or odor, turn off the device, and discontinue use. Contact Fisher Wallace Laboratories immediately. The long term corrective action by the company is still being determined. We request your action by replying to this email to confirm that you understand the information in this medical device correction and the subsequent actions needed to be taken. Fisher Wallace apologizes for any inconvenience this may cause. If you have any questions about what to do with your product, please contact Fisher Wallace by telephone Monday - Friday, 10:00 AM to 6:00 PM, Eastern Time at (800) 692-4380 or by email at postmarketsurveillance@fisherwallace.com.
Initiated
2023-04-06
Root cause
Nonconforming Material/Component
Distribution
US Nationwide. Global Distribution.
Recall number
Z-1600-2023
Recall status
Open, Classified
CDRH record ID
199353
FDA classified
2023-05-17
Recall event number
91982

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗