Z-1602-2020: Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000

Recall record Z-1602-2020 for Stryker Communications. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1602-2020
Company in FDA record
Stryker Communications
FDA record creation date
Mar 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

Source facts

Action
On October 11, 2019, a "Urgent Medical Device Correction" letter was sent to all affected customers via FedEx. In addition to explaining the recall, the recall letter as customers to take the following action: Please assure this notification letter is disseminated within your organization. A Stryker Representative will begin scheduling visits to customer facilities to perform inspections. This correction is expected to last one year. During the inspection, the Stryker Representative will perform a visual inspection on the lower portion of the Chromophare Ceiling Mounted Surgical Light Suspension. If a nonconforming unit is found, Stryker will replace the non-conforming Chromophare Ceiling Mounted Surgical Light Suspension with a new Chromophare Ceiling Mounted Surgical Light Suspension. In the meantime, the lights can continue to remain in full use. If you have any questions or notice any issues regarding this correction: " Stryker Technical Support at 866-841-5663 or comm.techservices@stryker.com " Nicole Kubat, Senior RA Compliance Specialist at 469-470-4318 or Nicole.Kubat@stryker.com " E-mail: CommunicationsRA1@stryker.com
Initiated
2019-10-11
Root cause
Other
Terminated
2022-10-18
Distribution
US: Nationwide OUS: Canada
Recall number
Z-1602-2020
Recall status
Terminated
CDRH record ID
180047
FDA record created
2020-03-27
Recall event number
85083

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗