Skip to content

Public FDA record

Z-1602-2023: YelloPort Elite, REF: EO0509505, 5x95 Optical Trocar, Rx Only, SterileR

Recall record Z-1602-2023 for Surgical Innovations Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1602-2023
Company in FDA record
Surgical Innovations Ltd
FDA classification date
May 17, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Optical Trocars sterility assurance can not be guaranteed

Source facts

Action
On March 29, 2023, Surgical Innovations issued a "Urgent Field Safety Notice" to affected consignees via E-Mail. An update to the communication was sent on 4/17/2023 to affected consignees via E-Mail. In addition to informing consignees about the recall, Surgical Innovations ask consignees to take the following action: 1. Distributor should confirm receipt of this Field Safety Notice 2. Identify the devices held in stock at the distributor and quarantine affected batches awaiting further instruction. 3. Issue this Field Safety Notice to all end users (Hospital) that the distributor has supplied 4. All product still in stock at the end users must be returned to the distributor. 5. Confirmation must be obtained by the distributor that either all stock has been returned or that any other product has been successfully used. 6. Confirmation to Surgical Innovations using the attached form of this disposition of all stock 7. Await further instruction from Surgical Innovations regarding further action for the affected products.
Initiated
2023-03-28
Root cause
Other
Distribution
US: MA OUS: None
Recall number
Z-1602-2023
Recall status
Open, Classified
CDRH record ID
199713
FDA classified
2023-05-17
Recall event number
92114

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗