Z-1602-2024: Alinity m System, Part Number: 08N53-002

Recall record Z-1602-2024 for Abbott Molecular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1602-2024
Company in FDA record
Abbott Molecular, Inc.
FDA classification date
Apr 19, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.

Source facts

Action
Abbott notified consignees on about 03/20/2024 via FedEx letter. Consignees were instructed to ensure the sample tube label requirements. If error code 6013 occurs and uncapped sample tubes are present on the sample rack, one of the following options can be used to resolve the error: 1. Aliquot the sample into another tube permitted by the assay package insert and ensure the sample tube label requirements in Appendix A are met; or 2. Cover colored portion of tube label that is present in the Keep-Out Zone with white opaque material (e.g. white sticker) where the barcode is located. Avoid covering any needed sample information when applying an additional sticker to the sample tube. Or 3. Reach out to the manufacturer of the tube that is triggering the error to discuss alternative tube types that meet the specifications documented within the Abbott assay specific package insert and do not have color within the required color free space (Keep-Out Zone). Consignees were also instructed to review this with appropriate personnel and consignees if further distributed, and to complete and return the response form.
Initiated
2024-03-20
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CAMEROON, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KENYA, LATVIA, LUXEMBOURG, MALAWI, MALAYSIA, MEXICO, MONTENEGRO, MOROCCO, MOZAMBIQUE, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PANAMA, PARAGUAY, POLAND, PORTUGAL, ROMANIA, RUSSIA (CIS), SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TURKEY, UGANDA, UNITED KINGDOM, VIETNAM, ZIMBABWE.
Recall number
Z-1602-2024
Recall status
Open, Classified
CDRH record ID
206738
FDA classified
2024-04-19
Recall event number
94251

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗