The upper deck may collapse to its lowest position and the caster brakes may fracture if the bed is moved while the brakes are locked.
Source facts
Action
Consignees were notified via letter dated April 8, 2008 to lower these beds to their lowest position, to remove the beds from use if they move more than 1/2 inch with the brakes locked and that the firm will repair these beds when fixes become available.
If you have questions regarding this recall, contact Hill-Rom Technical Support at 800-445-3730.
Initiated
2008-04-08
Root cause
Device Design
Terminated
2010-10-13
Distribution
Worldwide: USA and Canada.
Recall number
Z-1603-2008
Recall status
Terminated
CDRH record ID
69543
FDA classified
2008-09-05
Recall event number
47469