Skip to content

Public FDA record

Z-1603-2009: DERMALIFE Spa Feng Shui 2G. Part #E30007-100 is: Spa-Feng Shui 2G;100V 50/60Hz. This is the Spa Feng

Recall record Z-1603-2009 for Sybaritic, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1603-2009
Company in FDA record
Sybaritic, Inc
FDA classification date
Jul 13, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA. Sybaritic, Inc. conducted a recall on DERMALIFE 2G equipment (SpaJet, Spa Oceana, Spa Feng Shui) because of IR Fan failure which resulted in the overheating and failure of the IR emitter and surrounding assembly.

Source facts

Action
Sybaritic, Inc. issued a "Safety Alert" dated April 15, 2008 informing "All Owners/Users of DermaLife 2G Equipment" of the affected device. The letter advised consignees on device use until the mandatory Filed Correction Kit is delivered. For further questions, contact Sybaritic, Inc. at 1-800-445-8418.
Initiated
2007-06-22
Root cause
Device Design
Terminated
2012-07-05
Distribution
Worldwide Distribution -- US including states of CA, IA, MN, NE, NJ, NY, TX and WA and countries of Macedonia, Hungary, Greece, South Africa, Italy, Hong Kong, United Arab Emirates, Greece, Croatia, Australia, Kuwait, Turkey, Mexico, Belgium, England, Barbados, France, Japan, Lebanon, Moldova, Poland, Portugal, Romania, Russia, Spain and Thailand.
Recall number
Z-1603-2009
Recall status
Terminated
CDRH record ID
81947
FDA classified
2009-07-13
Recall event number
52036

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗