Z-1603-2024: IBA Proton Therapy System - PROTEUS 235, PTS-10 versions, PTS-11 versions, and PTS-12 versions.

Recall record Z-1603-2024 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1603-2024
Company in FDA record
Ion Beam Applications S.A.
FDA classification date
Apr 19, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.

Source facts

Action
On 3/20/2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers in person or via email by their customer representative. Customers were instructed to review the notice with their customer representative. Immediate actions taken: IBA carried out an analysis to determine if the impacted sites have some safety parameters checked by the BAPP in their clinical site configuration with values leading to violations. The result of this analysis showed that there are no sites with safety parameters checked by the BAPP out of tolerance. Intermediate action: While waiting for the solution to be deployed on your site, a regular automatic check that the safety parameters are within tolerances will be implemented. IBA will distribute an Internal User Notice to IBA operators of impacted sites to inform them that operators shall monitor alarms triggered by this check and directly inform you if a violation is detected. The intermediate action will be implemented for your site by April 30th, 2024, at the latest. The final solution will be implemented for your site by December 2025.By signing the response form, the customer representative confirms that this notice has been read, understood and communicated to the appropriate employees within the organization. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness of this notice and the resulting action for an appropriate period to ensure the effectiveness of the corrective action. The Food and Drug Administration has been informed of this Urgent Medical Device Correction. We apologize for any inconvenience that this may cause, and we would like to thank you for your cooperation. Your IBA representative can provide you with additional information and/or guidelines if necessary. Please contact Sonia Pinel, Head or Post Market Vigilance at +32 10 497 516 or Vigilance@iba-group.com or the helpdesk +32 2 507 20 81 (available 24/7).
Initiated
2024-03-04
Root cause
Device Design
Distribution
US Nationwide distribution in the states of AR, FL, LA, KS, MI, TX, VA.
Recall number
Z-1603-2024
Recall status
Open, Classified
CDRH record ID
206878
FDA classified
2024-04-19
Recall event number
94297

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗