Skip to content

Public FDA record

Z-1604-2020: DeRoyal SECURITY CUFFS, Single Strap, QR Buckle Cuff and Bed, UNIVERSAL, REF M8138 - Product Usage: patie

Recall record Z-1604-2020 for DeRoyal Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1604-2020
Company in FDA record
DeRoyal Industries Inc
FDA record creation date
Mar 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One lot of product was inadvertently manufactured with a buckle which does not meet the mechanical hold force required for this device.

Source facts

Action
DeRoyal issued their recall on 2/27/2020 to the end user level by emailing its direct accounts and sales representatives. The customers were asked to identify the affected products, quarantine them, return the affected product and the provided notice of return form back to DeRoyal indicating if replacement or credit is requested. Distributors were asked to further notify their customers. They were asked to do this directly or to provide the recalling firm with a customer listing and the recalling firm would notify them.
Initiated
2020-02-27
Root cause
Nonconforming Material/Component
Terminated
2022-05-10
Distribution
VA, CA, FL, AZ
Recall number
Z-1604-2020
Recall status
Terminated
CDRH record ID
180091
FDA record created
2020-03-30
Recall event number
85102

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗